AVE-733A

AVE-733A

This urine analyzer operates on reflectance photometry and measures a wide range of parameters including blood, leucocytes, glucose, proteins, and kidney function indicators like microalbumin and creatinine. It offers high throughput of up to 514 tests per hour, with automatic error correction and self-diagnostics to ensure reliable results. The device features a touch screen interface, automatic strip handling, built-in and external printing options, and large storage capacity for over 2 million test results. Compact and lightweight, it maintains accuracy and stability under typical lab conditions while supporting both clinical and research needs.

For more info (registered only): Dowlnoad Datasheet

Test Principle

  • Reflectance photometry

Detected Parameters

  • Occult blood, leucocytes, specific gravity, pH, protein, urobilinogen, bilirubin, ketones, glucose, nitrite, ascorbic acid, microalbumin, creatinine, calcium

  • Albumin/creatinine ratio (ACR) reportable

Features

  • High throughput (up to 514 tests/hour)

  • Automatic correction for non-specificity, pH, SG, and color interference

  • LCD touch screen, user-friendly interface

  • Automatic collection of used test strips

  • Built-in low-noise thermal printer, external printer support

  • Automatic saving/uploading of results

  • Power-on self-check, automatic fault reporting and error warning

  • Original NMPA-approved quality control materials available

Specifications

  • Throughput: ~514 tests/hour

  • Test strips: 10–14 parameters optional

  • Repeatability: CV ~1%

  • Accuracy: within one level of target values; no false negatives/positives with reference solutions

  • Stability: CV ~1% with 8 hours of continuous power

  • Storage: up to 2,100,000 results

  • Environmental: 30–80% relative humidity, 5–40 °C operating temp

  • Power: AC 100–240V, 50/60Hz, 250VA

  • Dimensions: 380 × 320 × 190 mm

Weight: 4.5 kg